Recall Z-1651-2013 ยท MZA Test, Equipment, Automated Bloodborne Pathogen

Ortho-Clinical Diagnostics

ORTHO Assay Software (OAS) Server, Catalog # 6842840 Usage: The ORTHO Summit System (OSS) is a modular, integrated system for use with licensed blood screening tests. Consisting of networked processors and instruments linked together via a local area network, assay-specific pipetting protocols and ORTHO Assay Software (OAS), OSS automates many of the processing functions and data management requi

DateJul 2, 2013 (initiated May 6, 2013)
StatusTerminated
ReasonThere is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.
Root causeDevice Design
ActionOrtho Clinical Diagnostics (OCD) sent an Urgent Product Correction Notification letters dated May 6, 2013 and response forms to all affected customers. Ortho Clinical Diagnostics informed their customers of the issue and advised if they detect the smell of melting plastic or observe smoke originating from your VITROS System, to immediately power down the system and contact OCD for assistance. They were also advised to consider performing the backup procedure more frequently to help prevent the loss of data on their Systems. In the near future, an OCD Field Engineer will contact customers who
Quantity31 units
DistributionWorldwide Distribution - USA Nationwide including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI & WV, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, J
FDA recordOpen on fda.gov

510(k) clearances named in this recall

BK090074Not in our 510(k) copy
BK130006Not in our 510(k) copy