Recall Z-1537-2013 ยท MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator

Respironics California Inc

Respironics V60 Ventilator, Model #V60 Respironics Material P/N (Philips 12 Digit P/N): 85008 (85008) 1053613 (989805628251) 1053614 (989805612101) 1053615 (989805613391) 1053616 (989805613661) 1053617 (989805611761) 1053618 (n/a) R1053618 (n/a) 1076709 (n/a) 1076715 (989805627411) 1076716 (989805627431) 1076717 (989805627441) DU1053617 (989805616411) U1053614 (989805636441) U1053617 (989805636

DateJun 17, 2013 (initiated Jun 4, 2013)
StatusTerminated
ReasonRespironics California, Inc. has initiated a recall on the V60 ventilator because of an issue with the software on the V60 Power Management Board Assembly. If a component fails on the Power Management Board Assembly, it may cause ventilator support to be lost with potentially no audible alarm from the ventilator.
Root causeSoftware design
ActionThe firm, Philips Healthcare, sent a "MEDICAL DEVICE CORRECTION" letter dated June 3, 2013 to all customers who received the V60 Ventilators. The letter describes the product, problem identified and the actions to be taken. The customers were informed that the V60 ventilator may continue to be used in accordance with its directions for use, pending the completion of the update. In addition, the customers were instructed to refer to their manual for additional information on warnings. A Philips Field Service Engineer, Approved Sevice Provider or Distributor will be contacting the customers t
Quantity19,200 units total (12,060 units in US)
DistributionWorldwide distribution: US (nationwide) and countries of: Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt , Finland, France, Germany, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel,