Recall Z-1929-2013 ยท MBL Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous
Smith & Nephew Inc
R3 (TM) XLPE ACETABULAR LINER UHMWPE, 20o OVERHANG, 36 MM ID, 60 MM OD, QTY: (1), STERILE H2O2, REF 71335760. Orthopaedic.
| Date | Aug 8, 2013 (initiated May 29, 2013) |
| Status | Terminated |
| Reason | One batch of 60 mm 20o R3 XLPE liner may be laser etched and packaged as a 62 mm 20o liner and one batch of 62 mm 20o R3 XLPE liner may be laser etched and packaged as a 60 mm 20o liner. |
| Root cause | Packaging process control |
| Action | All affected Smith & Nephew Sales Representatives were notified of problem and the recall via email on 5/29. They were instructed to cease use and distribution of the product and quarantine for return. Surgeons were notified by letter beginning on 6/12. International consignees were notified on 05/29 via email. |
| Quantity | 53 units |
| Distribution | Worldwide Distribution-USA including the states of PA, WA, TX, FL, IA, MI, WI, TN, KY, IN, NY, MO and OR, and the countries of India, Canada, Chile, Venezuela, Dubai, South Africa, Australia, Poland, Romania, Belgium, Germany and United Kingdom. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K113848 | R3 XLPE LINERS Smith & Nephew, Inc., Apr 27, 2012 |