ClearanceBrief
Recall Z-1568-2013 ยท KYI Prosthesis, Wrist, Carpal Trapezium

Integra LifeSciences Corp.

CMC Saddle Product Usage: The Ascension CMC is intended to resurface the proximal end of the first metacarpal in cases of rheumatoid arthritis, traumatic arthritis, osteo-arthritis, or post-fracture deformation or bone loss which present as either a painful, unstable thumb, or a thumb with limited range of motion. This is an uncemented prosthesis designed for press fit use only.

DateJun 21, 2013 (initiated Jun 3, 2013)
StatusTerminated
ReasonAs a result of a quality review, labelling content discrepancies were identified between the Instruction for Use (IFU) distributed in the US with the Ascension CMC and the content that is cleared by FDA.
Root causeLabeling Change Control
ActionIntegra sent an Urgent: Medical Device Recall letter to all affected product. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory, discontinue use, and complete and return the attached form. When the form is received, Customer Service Returns and Repairs will send an RMA number to return the affected inventory. For quwstions contact your regional managers or call 512-520-7697.
Quantity467 devices
DistributionUSA Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061451ASCENSION, MODELS CMC-400,13,CMC-400-20,CMC-400-30,CMC-400-40
Ascension Orthopedics, Inc., Aug 11, 2006