ClearanceBrief
Recall Z-1744-2013 ยท IBJ Light, Microscope

Zeiss, Carl Inc

Polarizer D Fixed w/Rotating Red 1 lambda plate. Utilized in upright Zeiss microscopes to aid in diagnosis by means of color variation.

DateJul 17, 2013 (initiated May 28, 2013)
StatusTerminated
ReasonCarl Zeiss Microscopy, LLC is recalling the Polarizer D component used in All upright ZEISS microscopes (Axio Imager, Axio Scope.A1, Axio Lab.A1) due to Misaligned Full-Wave (Lambda) Plate.
Root causeNonconforming Material/Component
ActionOn May 28, 2013, Carl Zeiss Microscopy, LLC recalled a Polarizer Component due to Misaligned Full-Wave (Lambda) Plate. Carl Ziess Microscopy distributed notices to their customers via e-mail. Carl Zeiss Microscopy, LLC requests that all customers immediately discontinue the use of the component for any applications which rely on accurate color-change sequence results. Your service provider will contact you to schedule a visit in which a factory trained service technician will confirm the alignment of your Polarizer D component;s lambda plate. If it is defective, a new Polarizer D component
Quantity10
DistributionDistributed in the states of MI, TX, SD, FL, GA, NY, and NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.