Recall Z-0329-2014 ยท FEQ Pump, Air, Non-Manual, For Endoscope

Olympus America Inc.

Olympus Endoscopic Flushing Pump, Model OPF-2. Facilitates washing of tissues to remove blood, feces and other organic matter during endoscopic procedures.

DateNov 15, 2013 (initiated May 20, 2013)
StatusTerminated
ReasonThere is a possibility that the pump head was not shipped in the locked position. If the pump head is in an unlocked position this may, under certain circumstances, cause the pump to stop and go into power stand-by mode.
Root causeComponent design/selection
ActionNotification letters were sent to customers on May 20, 2013.
Quantity263
DistributionWorldwide distribution: USA (nationwide) and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100899OLYMPUS ENDOSCOPIC FLUSHING PUMP, MODEL OPF-2
Keymed (Medical & Industrial Equipment), Ltd., Sep 13, 2010