ClearanceBrief
Recall Z-1495-2013 ยท FYA Gown, Surgical

Stryker Instruments Div. of Stryker Corporation

Stryker Instruments 0400-810-000 T4 Zipper Toga (S/M) Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.

DateJun 6, 2013 (initiated May 23, 2013)
StatusTerminated
ReasonBased on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr
Root causeNonconforming Material/Component
ActionStryker sent an Urgent Urgent Recall noitifications letter dated May 7, 2013 via certified US mail and E-mail l to all affected customers, distributors and direct accounts. The letter identified the affected products, problem, risk to health, and actions to be taken. Customers were instructed to check inventory, immediately quarantine any affected product, and open each package before discarding to prevent use. Customers were directed to fill out and return a Business Reply Form to Stericycle, Phone: 888-943-2396 Fax: 888-943-4192, Email: stryker5893@stericycle.com verifying directions an
Quantity7960 units (796 boxes)
DistributionUSA Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070078STRYKER T4 HYTREL TOGA
Stryker Instruments, Instruments Div., Feb 2, 2007