Recall Z-1641-2013 ยท NBH Accessories, Arthroscopic
Surgical Instrument Service And Savings, Inc.
- CLEAR-TRAC COMPLETE, Threaded Cannula with Obturator, (Grey) 7.0mm x 72mm Ref. 72200905 Lot 122631, 122609 - CLEAR-TRAC Flexible Cannula System Threaded, w/Obturator, (Lime Green) 8.0mm x 72mm Ref. 72200425 Lot 122631 - CLEAR-TRAC Flexible Cannula, Threaded, w/Obturator, (Orange) 6.5mm x 72mm Ref. 72200427 Lot 122465 - Crystal Cannula w/Obturator (Orange) 5.75mm I.D. X 7cm Ref. AR-6460 Lot
| Date | Jul 1, 2013 (initiated May 3, 2013) |
| Status | Terminated |
| Reason | The seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch. |
| Root cause | Packaging process control |
| Action | MEDISISS sent a Correction Immediate Action Request letter dated May 2, 2013, to all affected customes. The letter identified the product, the problem, and the action to be taken by the customer. The letter directed the customer to inspect the products prior to use for seal integrity as specified on the product labeling: "Sterile unless package has been damaged or opened. Do not use if package is damaged." This will be followed up by a visit from the customer's sales representatives who are tasked with inspecting all the lots that may be remaining in the customer's inventory. The sales r |
| Quantity | 73 |
| Distribution | Nationwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K012667 | REPROCESSED ARTHROSCOPIC SHAPERS Surgical Instruments Service and Savings, Inc., Jun 3, 2002 |