Recall Z-1638-2013 ยท JOW Sleeve, Limb, Compressible

Surgical Instrument Service And Savings, Inc.

- Calf Garment Standard Ref. L501-M Lot 122549, 122985 - Calf Garment, DVT10 Standard Ref. DVT10 Lot 122570, 12546 - Calf Garment, DVT20 Large Ref. DVT20 Lot 122549, 122570 - SCD Express Compression Sleeve, Knee Length, (White) Large Ref. 9789 Lot 122574 - SCD Express Compression Sleeve, Knee Length, (White) Medium Ref. 9529 Lot 122547, 122574 - SCD Express Compression Sleeve, Thigh Length

DateJul 1, 2013 (initiated May 3, 2013)
StatusTerminated
ReasonThe seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch.
Root causePackaging process control
ActionMEDISISS sent a Correction Immediate Action Request letter dated May 2, 2013, to all affected customes. The letter identified the product, the problem, and the action to be taken by the customer. The letter directed the customer to inspect the products prior to use for seal integrity as specified on the product labeling: "Sterile unless package has been damaged or opened. Do not use if package is damaged." This will be followed up by a visit from the customer's sales representatives who are tasked with inspecting all the lots that may be remaining in the customer's inventory. The sales r
Quantity229
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012658REPROCESSED COMPRESSION LIMB SLEEVES
Surgical Instruments Service and Savings, Inc., May 14, 2002