Recall Z-1637-2013 ยท HWX Tap, Bone
Surgical Instrument Service And Savings, Inc.
TPS Micro Reciprocating Small Tear Cross Cut Rasp 5.0mm x 11mm Ref. 5100-37-113 Lot 122518, 122447. Orthopedic manual surgical instrument
| Date | Jul 1, 2013 (initiated May 3, 2013) |
| Status | Terminated |
| Reason | The seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch. |
| Root cause | Packaging process control |
| Action | MEDISISS sent a Correction Immediate Action Request letter dated May 2, 2013, to all affected customes. The letter identified the product, the problem, and the action to be taken by the customer. The letter directed the customer to inspect the products prior to use for seal integrity as specified on the product labeling: "Sterile unless package has been damaged or opened. Do not use if package is damaged." This will be followed up by a visit from the customer's sales representatives who are tasked with inspecting all the lots that may be remaining in the customer's inventory. The sales r |
| Quantity | 4 |
| Distribution | Nationwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.