Recall Z-1633-2013 ยท HRX Arthroscope
Surgical Instrument Service And Savings, Inc.
- Abrader Bur, Round (Aqua) 4.0mm Ref. 7205324 Lot 123270 - AcromioBlaster Burr (Brick Red) 5.5mm Ref. 7205669 Lot 122590, 122454, 122465, 122434, 122665 - AcromioBlaster Burr (Sage Green) 4.0mm Ref. 7205668 Lot 122590, 123270 - Acromionizer Bur, Barrel, (Mauve) 4.0mm Ref. 7205326 Lot 122609 - BoneCutter Blade, Full Radius, Series 3000 (Orange/Black) 5.5mm Ref. 7206010 Lot 122609 - CoolCut Bo
| Date | Jul 1, 2013 (initiated May 3, 2013) |
| Status | Terminated |
| Reason | The seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch. |
| Root cause | Packaging process control |
| Action | MEDISISS sent a Correction Immediate Action Request letter dated May 2, 2013, to all affected customes. The letter identified the product, the problem, and the action to be taken by the customer. The letter directed the customer to inspect the products prior to use for seal integrity as specified on the product labeling: "Sterile unless package has been damaged or opened. Do not use if package is damaged." This will be followed up by a visit from the customer's sales representatives who are tasked with inspecting all the lots that may be remaining in the customer's inventory. The sales r |
| Quantity | 907 |
| Distribution | Nationwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K012667 | REPROCESSED ARTHROSCOPIC SHAPERS Surgical Instruments Service and Savings, Inc., Jun 3, 2002 |