ClearanceBrief
Recall Z-1629-2013 ยท GEN Forceps, General & Plastic Surgery

Surgical Instrument Service And Savings, Inc.

- Re-New II Endocut Scissor Tip, 16.51mm 5.0mm Ref. 3142 Lot 122295, 122308, 122455, 122875 Manual surgical instrument for general use.

DateJul 1, 2013 (initiated May 3, 2013)
StatusTerminated
ReasonThe seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch.
Root causePackaging process control
ActionMEDISISS sent a Correction Immediate Action Request letter dated May 2, 2013, to all affected customes. The letter identified the product, the problem, and the action to be taken by the customer. The letter directed the customer to inspect the products prior to use for seal integrity as specified on the product labeling: "Sterile unless package has been damaged or opened. Do not use if package is damaged." This will be followed up by a visit from the customer's sales representatives who are tasked with inspecting all the lots that may be remaining in the customer's inventory. The sales r
Quantity10
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.