Recall Z-1583-2013 ยท GJS Test, Time, Prothrombin

Roche Diagnostics Operations, Inc.

CoaguChek XS Plus System The CoaguChek XS Plus system for use by professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy.

DateJun 24, 2013 (initiated May 16, 2013)
StatusTerminated
ReasonRoche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.
Root causeSoftware design
ActionRochet sent an Urgent Medical Device Correction letter dated May 16, 2013, via UPS Ground to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were issued the following instructions: If you are using Warfarin therapy in combination with antibiotics and/or chemotherapy drugs, which could potentially lead to an extremely high test result (> 10 INR), you may in rare cases receive repeated ERROR 6 messages instead of a test result. If the ERROR 6 message is displayed repeatedly on your CoaguChek XS meter, immediately
Quantity3343 meters
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071041COAGUCHEK XS PLUS SYSTEM
Roche Diagnostics, May 11, 2007