ClearanceBrief
Recall Z-1581-2013 ยท GJS Test, Time, Prothrombin

Roche Diagnostics Operations, Inc.

CoaguChek XS System (PST) The CoaguChek XS PT test strips are part of the CoaguChek XS System. The CoaguChek XS System measures blood-clotting time for people who are taking anticoagulation medications such as Coumadin or warfarin.

DateJun 24, 2013 (initiated May 16, 2013)
StatusTerminated
ReasonRoche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.
Root causeSoftware design
ActionRochet sent an Urgent Medical Device Correction letter dated May 16, 2013, via UPS Ground to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were issued the following instructions: If you are using Warfarin therapy in combination with antibiotics and/or chemotherapy drugs, which could potentially lead to an extremely high test result (> 10 INR), you may in rare cases receive repeated ERROR 6 messages instead of a test result. If the ERROR 6 message is displayed repeatedly on your CoaguChek XS meter, immediately
Quantity204,205 meters total
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062925COAGUCHEK XS SYSTEM
Roche Diagnostics Corp., Jan 29, 2007