ClearanceBrief
Recall Z-1455-2013 ยท MEB Pump, Infusion, Elastomeric

Symbios Medical Products, LLC

Part 510121 Disposable Infusion Pump, ...Elastomeric Pump 300 mL, 4mL/hr,...sterile.. found in the following kits GoPump Kit 510203-BP, Epidural catheter & BIOPATCH GoPump Kit 510204, 2.5" Fenestrated catheter GoPump Kit 510204 BP, 2.5 Fenestrated catheter & Biopatch GoPump Kit 510205BP, 5" Fenestrated catheter & Biopatch GoPump Kit 510551-BP, 2.5 Fenestrated catheter & Biopatch GoBlock kit 510

DateJun 6, 2013 (initiated May 10, 2013)
StatusTerminated
ReasonSymbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the
Root causeUnder Investigation by firm
ActionSymbios sent UrgentMedical Device Recall notices to all consignees beginning May 10, 2013. A second corrected/amended notice was sent on May 14, 2013 to consignees and a National Press release was issued. May 30,2013, a third amended Urgent Recall Notice was sent to all consignees correcting the pump description and lot number range table on page 5 of the notice. The notices and press release provided identification of the recalled devices including devices in the previous February 2013 recall, described possible adverse events, and included directions to to quarantine product in a secure
Quantity33,543 in all kits
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.