Recall Z-1625-2013 ยท GAG Stapler, Surgical
Surgical Instrument Service And Savings, Inc.
-ENDOPATH ETS-Flex Articulating Endoscopic Linear Cutter, (Blue) w/o Staples 35mm. REF ATB35 Lot 122300. -Ligaclip MCA Multiple Clip Applier, 20 Small Clips Small. REF MCS20 Lot 122300. - Ethicon Pistol Grip, Fixed Head Skin Stapler, Multi-Directional Release, (Grey) 35W. REF PXW35 Lot 122309 and Lot 123263. -Proximate Linear Cutter, w/o Staples 75mm. REF TLC75 Lot 122490 -Ethicon Proximat
| Date | Jul 1, 2013 (initiated May 3, 2013) |
| Status | Terminated |
| Reason | The seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch. |
| Root cause | Packaging process control |
| Action | MEDISISS sent a Correction Immediate Action Request letter dated May 2, 2013, to all affected customes. The letter identified the product, the problem, and the action to be taken by the customer. The letter directed the customer to inspect the products prior to use for seal integrity as specified on the product labeling: "Sterile unless package has been damaged or opened. Do not use if package is damaged." This will be followed up by a visit from the customer's sales representatives who are tasked with inspecting all the lots that may be remaining in the customer's inventory. The sales r |
| Quantity | 13 |
| Distribution | Nationwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.