Recall Z-1433-2013 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

8mm Fenestrated Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.

DateMay 29, 2013 (initiated May 8, 2013)
StatusTerminated
ReasonReports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.
Root causeDevice Design
ActionIntuitive Surgical sent an "Urgent Medical Device Correction" letter dated May 8, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Customer Service at 800-876-1310 for questions regarding this notice.
Quantity697 - all products
DistributionWorldwide Distribution-USA (nationwide) including the states of AL, CA, CO, FL, IL, KY, MI, MS, NJ, NY, PA, TX, UT, VA, WA, and WI, and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Netherlands, Norway, Qatar, Romania, Spain, Sweden, Switzerla
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012833INTUITIVE SURGICAL BIPOLAR FORCEPS
Intuitive Surgical, Inc., Nov 16, 2001