Recall Z-1525-2013 ยท LTT Panels, Test, Susceptibility, Antimicrobial

Siemens Healthcare Diagnostics, Inc.

Siemens brand MicroScan Dried Negative Breakpoint Combo 47 panels (B1017-417), SMN #: 10483099), For use in determining quantitative and/or qualitative antimicrobial agent susceptibility.

DateJun 8, 2013 (initiated May 2, 2013)
StatusTerminated
ReasonThe product has the potential for false negative or delayed reaction for Arginine (ARG) results. In addition, 42 hour identification panel holds may occur due to a weak ARG reaction.
Root causeNonconforming Material/Component
ActionSiemens sent an Urgent Field Safety Notice dated May 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact your local Siemens Healthcare Diagnostics Representative for information concerning this notice.
Quantity683 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.