ClearanceBrief
Recall Z-1507-2013 ยท FGE Stents, Drains And Dilators For The Biliary Ducts

Cordis Corporation

FLEXSTENT Biliary Self-Expanding Stent System. Product Usage: The FSS FLEXSTENT Biliary SES product is indicated for use in the palliation of malignant strictures in the biliary tree.

DateJun 6, 2013 (initiated Apr 23, 2013)
StatusTerminated
ReasonHoles and/or tears can be introduced in the FLEXSTENT inner pouch under some conditions, and to a lesser degree, in the outer pouch.
Root causePackage design/selection
ActionCordis sent an Urgent Medical Device Correction and Potential Removal letter dated April 24, 2013 e-mail to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately identify and set aside all affected products to prevent further use of the affected product, until inspected. Use only product identified with a blue dot next to the lot number on the carton face label, carton end label, and outer pouch label, or product with upgraded sterile barrier system (not yet available, as of the date of this letter), rev
Quantity23 units
DistributionWorldwide Distribution - USA (nationwide) and the states of: AZ, CA, CT, FL, GA, IA, IL, LA, MD, MI, MS, NC, NJ, NY, OH, PA, RI, SC, TX, WA, WI, WV, and the countries of: Czech Republic, Italy, Netherlands, and New Zealand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081996FLEXSTENT BILIARY SELF EXPANDING STENT SYSTEM
Flexible Stenting Solutions, Inc., Sep 15, 2009