Recall Z-0861-2014 ยท JXG Shunt, Central Nervous System And Components
Codman & Shurtleff, Inc.
Codman Certas Programmable Valve In Line Valve only with SiphonGuard Device Product Code: 82-8804 Product Usage: The CODMAN CERTAS Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF (cerebrospinal fluid) for the management of hydrocephalus. The valve can be set to a choice of eight opening pressure settings for constant intraventricula
| Date | Jan 29, 2014 (initiated May 16, 2013) |
| Status | Terminated |
| Reason | Codman Certas Programmable Valves used for hydrocephalus may not operate properly |
| Root cause | Device Design |
| Action | Codman issued an Urgent Notice dated May 16, 2013, to all affected customers. The letter identifed the affected product, problem and actions to be taken. Mitigation instructions are provided and if not followed or are not successful, either issue could lead to symptoms associated with under- or over-drainage. Units not implanted are to be returned. A U.S Acknowledgement Form and fax are requested to be completed and faxed to: 1-888-943-4193, or e-mail the form to codman8894@stericycle.com. If you have any questions on concerns in regards to this recall, please contact your Codman Repres |
| Quantity | 1481 units |
| Distribution | Worldwide Distribution - US Nationwide, Australia, Austria, Belgium, Brazil, Switzerland, Czech Republic, Germany, Spain, Finland, France, Great Britain, Israel, Italy, Netherlands, Norway, Portugal, Russia, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa, and UK. |
| FDA record | Open on fda.gov |