Recall Z-1753-2013 ยท OWB Interventional Fluoroscopic X-Ray System

Leibel-Flarsheim Company LLC

Hydra Vision DR 60/80 System, Models 700539 and 700540, Liebel-Flarshiem Company. The System is used in radiologic imaging procedures.

DateJul 18, 2013 (initiated Apr 25, 2013)
StatusTerminated
ReasonA complaint was received in which a hospital physicist indicated the dosage was too high on the HUT system. During the service call the systems Maximum Fluoro Dose was found to be 30 R/min which is three times the Maximum Fluoro Dose limit of 10 R/min.
Root causeEmployee error
ActionThe customers received an "Urgent Device Correction" notification letter on May 12, 2013 explaining the product issue and providing recommended actions. Customers were contacted to set up a service call to test and recalibrate equipment as necessary. Questions were directed to 800-778-7898.
Quantity12 units
DistributionNationwide Distribution (US), including the states of TN, KY, CO, KS, CT, HI, and AZ.
FDA recordOpen on fda.gov