Recall Z-1727-2013 ยท JMM Column Chromatography & Color Development, Hydroxyproline

Roche Diagnostics Operations, Inc.

B-CrossLaps/serum (B-CTx in serum) REF 11972308 160 100 tests Immunoassay for the in vitro quantitative determination of degradation products of type I collagen in human serum and plasma as an aid in assessing bone resorption. The test may be used as an aid in monitoring antiresorptive therapies (e.g. bisphosphonates, hormone replacement therapy - HRT) in postmenopausal women and individuals dia

DateJul 12, 2013 (initiated Apr 25, 2013)
StatusTerminated
ReasonRoche Diagnostics internal investigations in R&D have shown that the claims for biotin interference and the sample stability given in the Elecsys -CrossLaps (-CL) assay package insert version 5, could not be reproduced. The -CrossLaps test results contribute to the monitoring of antiresorptive therapy. False low or false normal results may lead to delayed diagnosis and treatment of osteoporosi
Root causeDevice Design
ActionThe firm, Roche, sent an "URGENT MEDICAL DEVICE CORRECTION" (UMDC)letter dated May 2, 2013, to its consignees/customers via UPS Ground. The letter describes the product, problem and actions to be taken. The consignees/customers were instructed to apply the correct sample storage conditions as stated in the UMDC and to inform all relevant parties about the correct biotin interference claim. The consignees/customers were instructed to keep the UMDC on file for reference until the revised package insert is available. The consignees/customers were asked to complete and return a form via fax (
Quantity3056 units
DistributionNationwide distribution: USA including states of: AL, CA, GA, FL, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NY, OH, PA, TX, VA, VT, and WA.