Recall Z-1943-2013 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare, LLC

SIGNA INFINITY MR SYSTEM WITH EXCITE TECHNOLOGY GE 3.0T SIGNA INFINITY TWINSPEED WITH EXCITE MR SYSTEM GE SIGNA 3.0T WITH EXCITE MR SYSTEM GE SIGNA EXCITE 1.5T MR SYSTEM GE SIGNA EXCITE 3.0T MR SYSTEM Signa HDx Signa HDxt Signa HDi Signa Vibrant GE 0.35T SIGNA OVATION WITH EXCITE MR SYSTEM (i.e. Signa Ovation 5 and Ovation HD) GE 1.5T SIGNA HDe MR SYSTEM GE 0.35T SIGNA OVATION WITH EXCITE

DateAug 9, 2013 (initiated Nov 21, 2011)
StatusTerminated
ReasonScreen save images are included in wrong exam due to duplicate UID creation. This may cause a safety issue due to a potential patient data mismatch.While the issue has only been observed at one customer site on the Signa HDxT 1.5T system, internal investigation revealed that the issue also impacts multiple GE healthcare MR Systems.
Root causeSoftware design
ActionGE Healthcare communicated internally to their Service Team the directions for completing the correction on the customers' systems. GE Healthcare's recall strategy was to update all affected systems that were distributed. At this time,Week of 02/06/2012, deployment of this recall is considered complete. Should you have any questions or require additional information, please contact Regulatory Affairs Leader by telephone at 262-548-2608 or email paulcorrigan@ge.com.
Quantity7016 (2919 US; 4097 OUS)
DistributionWorldwide distribution: USA (nationwide) including DC and Puerto Rico and countries: ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIDJAN, BELARUS, BELGIUM, BOLIVIA, BOSNIA-HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN RE
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013636SIGNA INFINITY MR SYSTEM WITH EXCITE TECHNOLOGY
GE Medical Systems, Jan 24, 2002
K030874GE 3.0T SIGNA INFINITY TWINSPEED WITH EXCITE MR SYSTEM
General Electric Co., Mar 28, 2003
K040444GE SIGNA 3.0T WITH EXCITE MR SYSTEM
General Electric Co., Apr 14, 2004
K041476GE SIGNA EXCITE 1.5T MR SYSTEM, GE SIGNA EXCITE 3.0T MR SYSTEM
Ge Medical Systems, LLC, Jun 17, 2004
K052293GE 1.5T AND 3.0T SIGNA HDX MR SYSTEM
Ge Medical Systems, LLC, Sep 21, 2005