Recall Z-1558-2013 ยท CCW Laryngoscope, Rigid
Verathon, Inc.
GlideScope Video Laryngoscope (GVL) 3, 4, 5 and GlideScope (AVL) 2, 3, 4 and 5 reusable blades Product Usage: The GlideScope Video Laryngoscopes are intended for use by qualified medical professionals to obtain a clear, unobstructed view of the vocal cords for medical procedures. The GlideScope Video Laryngoscopes system should only be used by individuals who have been trained and authorized by
| Date | Jun 20, 2013 (initiated May 8, 2013) |
| Status | Terminated |
| Reason | Verathon GlideScope Reusable GVL and AVL Blades are being recalled due to the potential risk of premature failure/breakage of the blade tip that may not be readily visible during routine inspection before or after intubation. |
| Root cause | Device Design |
| Action | Verathon issued an Urgent - Safety Alert and Medical Device Recall letter, dated May 10, 2013, to all affected customers The letter identified the affected product, problem and actions to be taken. The letter advises customers to discontinue the use of the specific serial numbers listed in the letter and asked that the affected product be returned to Verathon for replacement. The customers are also directed to fill out a return response form (included in the letter) regarding this recall. Consignees with questions are directed to contact their Verathon representative or Verathon Customer |
| Quantity | 4784 units worldwide (3,381 in the US and 1403 international) |
| Distribution | Worldwide distribution: USA (nationwide) including states of : AK, AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Washington DC, Guam, and countries of : Argent |
| FDA record | Open on fda.gov |