Recall Z-1536-2013 ยท HWC Screw, Fixation, Bone

OrthoPediatrics Corp

90 Degree Cannulated Infant Blade Plate 35mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is intended for fixation of long bone fractures and osteotomies in all pediatric subgroups (except neonates) and in small stature adults. Specific indications include: intertrochanteric derotation and varus osteotomies, femoral neck and pertrochanteric fractures, i

DateJun 12, 2013 (initiated Apr 19, 2013)
StatusTerminated
ReasonPost Market Surveillance of the 90 degree Cannulated Infant Blade Plate indicated that there have been five reports concerning bending of the plate intra-operatively.
Root causeDevice Design
ActionOrthoPediatrics sent an Urgent Voluntary Recall notifiation via telephone and email to all affected consignees. The letter instructed consignees to take possession of any affected devices and quarantine immediately. A Mandatory Reply Form is included with the notification letter to be completed and returned. Please contact Greg Teghtmeyer if you have any questions (574-268-6379).
Quantity130 devices were shipped
DistributionUSA Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110959ORTHOPEDIATRICS BLADE PLATE SYSTEM
OrthoPediatrics Corp., Aug 5, 2011