Recall Z-1598-2013 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Toshiba American Medical Systems Inc

TOSHIBA Kalare Diagnostic X-Ray System K110785

DateJul 10, 2013 (initiated Apr 3, 2013)
StatusTerminated
ReasonFirm initiated recall because when radiographic conditions with the small focus are set and the radiography ready switch is held down, the large focus is also heated after a few seconds or more. This occurs when the Automatic Exposure Function (AEC) function is used. As a result, a radiographic tube current exceeding the set value flows and a brighter image than expected is acquired.
Root causeComponent design/selection
ActionFirm initiated recall by sending recall notification letters along with a customer reply form on April 3, 2013 via USPS return mail receipt. A follow-up letter (with customer reply form) was sent on April 11, 2013 to inform customers that they should continue to use the AEC function as usual. The updated letter along with the customer reply form dated April 11, 2013 was also sent via USPS return mail receipt. The letter titled, "URGENT: MEDICAL DEVICE CORRECTION", notified customers of the reason for recall, product description under recall, corrective action, instructions to customer under
Quantity153 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110785KALARE DIAGNOSTIC X-RAY TABLE SYSTEM