ClearanceBrief
Recall Z-1326-2013 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Hologic, Inc.

Fluoroscan Mini C-arm InSight, software versions 1.x, 2.x, 3.x The device is a Mini C-arm Fluroscopic Imaging System designed to provide physicians with general fluoroscopic visualization of a patient, including but no limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.

DateMay 17, 2013 (initiated Apr 25, 2013)
StatusTerminated
ReasonThe audible alarm to alert the operator when cumulative irradiation time reaches five minutes is not accurate and the total cumulative irradiation time is not accurate.
Root causeSoftware design
ActionThe firm, Hologic, Inc. sent an "Urgent: Medical Device Recall" letter dated April 25, 2013, via FedEx to consignees/customers notifying them of the software defect and providing them with corrected user installable software version. The customers were instructed to complete and return the Enclosure to Customer Notification, Confirmation of Notification form via fax to Hologic at 866-652-8674 or mail the document to Hologic using the enclosed pre-addressed, pre-paid return envelope. Should you have any questions regarding this recall, require further information or assistance with the softw
Quantity287
DistributionWorldwide distribution: USA (nationwide) and countries of: DE, RU, AU, IT, GB, ES, FR, KR, ZA, CA, AR, LT, TW, CN, BE, AT, JP, QA, SA, SG, HK, SI, NL, and SE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051025FLUOROSCAN INSIGHT MINI C-ARM FLUOROSCOPIC IMAGING SYSTEM
Hologic, Inc., May 13, 2005