ClearanceBrief
Recall Z-1957-2013 ยท HXX Screwdriver

SpineFrontier, Inc.

Invue Solid Driver, Tipped, AO. Part Number: 11-70083. Allows the user to retain the screw when changing axis of insertion during fixation of spinal intervertebral.

DateAug 13, 2013 (initiated May 9, 2012)
StatusTerminated
ReasonThe nipple of the tipped driver broke off when surgeons were attempting to insert the screw.
Root causeDevice Design
ActionOn May 9, 2012 an email was sent to consignees requesting that the product be returned for servicing and that a replacement part would be sent.
Quantity9
DistributionDistributed in DC and the states of MA, and OK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121060SPINE FRONTIER INDUS ACP SYSTEM
Spinefrontier, Inc., Jul 3, 2012