Recall Z-1325-2013 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems, Inc. Oncology Systems

Varian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29, HCX. Product Usage: The Varian High Energy Accelerator is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

DateMay 15, 2013 (initiated May 31, 2011)
StatusTerminated
ReasonUnder certain conditions, photon beams in High Energy Clinacs may experience a gradual change in beam symmetry, potentially reaching an asymmetry of up to about 7% before interlock occurs. At maximum asymmetry, this may result in no more than about 3.5% dose deviation from expected at any point in the beam, and may result in minor injury to the patient or slightly decreased local disease control.
Root causeProcess design
ActionVarian sent a Customer Technical Bulletin to all affected customers. The bulletin identified the affected product, problem and actions to be taken. For questions, issues, or concerns contact Varian at 1-888-827-4265 or go to http://my.varian.com
Quantity3943 units
DistributionWorldwide distribution - USA (nationwide) including Puerto Rico, and countries of: Algeria, Angola, Australia, Austria, Bahrain, Bangladesh, Belarus, Belguim, Brazil, Bosnia and Hezregovina, Bulgaria, Canada, Chile, China, Columbia, Cyprus, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finla
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100890VARIAN HIGH ENERGY LINEAR ACCELERATOR
Varian Medical Systems, Jul 13, 2010