Recall Z-1737-2013 ยท MNI Orthosis, Spinal Pedicle Fixation

SpineFrontier, Inc.

KRD1 PedFuse Rods, Model: IM80100-13 Rev B Product Usage: Pedicle rods are cylindrical rods which act as a linking mechanism between pedicle screws. The rod sits with the pedicle screw tulip and they are locked into place with a set screw or locking cap. The rod may be placed along several pedicle screws. Component rods are used during implantation of pedicle screw system devices intended to pr

DateJul 16, 2013 (initiated Dec 30, 2011)
StatusTerminated
ReasonOne of the rods in each of two test constructs did not meet fatigue requirement as outlined in ASTM F1717-04 dynamic compression testing 5 million cycles.
Root causeVendor change control
ActionSpineFrontier issued a Recall Notification to the affected distributor via telephone. The distributor was instructed to return the affected product for a replacement product.
Quantity2 rods
DistributionUSA Nationwide Distribution including the state of TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092420KRD1 PEDICLE SCREW SYSTEM
Spinefrontier, Inc., Nov 19, 2009