ClearanceBrief
Recall Z-1657-2013 ยท NAV Graft, Bypass, Coronary Artery

The Anspach Effort, Inc.

***REF SC2100***eMax 2 Plus Console***SN G 10307653002*** Product Usage: The RPM of the Motor Handpiece may be less than 80,000 RPM (approximately 72,000 RPM) in the free running state when using the optional Hand Control. The low Motor Handpiece RPM would only be experienced in the Free running state (Hand Control fully depressed and no load on the cutting bur). This condition would not affect t

DateJul 2, 2013 (initiated Jun 24, 2011)
StatusTerminated
ReasonThe RPM of the Motor Handpiece may be less than 80,000 RPM.
Root causeProcess control
ActionAnspach Effort sent an "URGENT: Medical Device Correction" letter dated June 24, 2011 to their customers via mail. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete a reply form promptly confirming receipt of their letter and return it by fax to the number provided on the form. If any of the affected products were distributed to other services or facilities, please forward this information as appropriate. For questions contact Anspach Customer Support at 800-327-6887 or email customer.support@synthes.com.
Quantity61 units
DistributionWorldwide Distribution - USA Nationwide including the states of: AZ, CA, FL, KY, ME, MI, MO, OH, PA, TN, TX , VA, and the countries of: Australia, Egypt, Italy, Japan, Switzerland, Taiwan, and Turkey;
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.