Recall Z-2132-2013 ยท NDN Cement, Bone, Vertebroplasty

DePuy Spine, Inc.

CONFIDENCE SPINAL CEMENT SYSTEM 5CC KIT, Product Code: 2839-05-000 The CONFIDENCE Spinal Cement System is intended for percutaneous delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures.

DateSep 3, 2013 (initiated Apr 17, 2013)
StatusTerminated
ReasonDuring injection of cement, the water in the hydraulic pump leaks past the piston within the pump body resulting in the loss of pressure and inability to continue to inject cement.
Root causeProcess control
ActionDePut Synthes Spine sent an Urgent Voluntary Product Recall Notification letter dated April 15, 2013, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the con signee. Consignees were instructed to review their inventory and quarantine for return any products listed in the recall notice., complete the enclosed Business Reply Form, indicating level of inventory of the affected product and return the form to Depuy via fax to 888-943-4897or email to DePuy7862@stericycle.com and to return the product with the enclosed UPS label to Stericy
Quantity3896 units (Overall quantity distributed appears greater than quantities manufactured due to the distribution pathways)
DistributionWorldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WV, WY; Worldwide: Algeria, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, France, Germany, Hong Kong, I
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060300CONFIDENCE HIGH VISCOSITY BONE CEMENT
Disc-O-Tech Medical Technologies, Ltd., Jun 21, 2006