ClearanceBrief
Recall Z-1277-2013 ยท KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)

Olympus America Inc.

Olympus SurgMaster UES-40 electrosurgical unit ("UES-40") Product Usage: Designed for use in a medical facility under the supervision of a trained physician. It has been designed for general (open) and endoscopic surgery including urology, gynecology, respiratory and gastroenterology in conjunction with Olympus designed electrosurgical accessories, endoscopes (fiberscopes, videoscopes and rigid

DateMay 8, 2013 (initiated Nov 10, 2009)
StatusTerminated
ReasonOlympus initiated a software upgrade and enhancement in 2009 and completed the action in 2010. The upgrade and enhancement were initiated due to complaints of overheating, smoke vapors and failure to operate associated with the Olympus SurgMaster electrosurgical unit, model UES-40 ("UES-40") when used in conjunction with the Olympus WA22557C electrode. The complaint investigations revealed that
Root causeUse error
ActionOlympus America Inc sent a Field Corrective Action letter dated November 10, 2009 to all affected consignees. The letter identified the affected product, problem and actions to be taked. Consignees were directed to contact Gyrus at 888-524-7266 to schedule return of the UES-40 generator for the software upgrade. The letter also directed consignees to contact OAI Customer Service at 800-848-9024 if they needed a new copy of the UES-40 Instruction Manual. For questions call 484-896-5688.
Quantity608
DistributionNationwide Distribution - USA Nationwide including the states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, and WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030194XUES-41 ENDOSCOPIC ELECTROSURGICAL UNIT
The Olympus Optical Co., Mar 24, 2003