Recall Z-1706-2017 ยท IZL System, X-Ray, Mobile

Philips Healthcare Inc.

Mobile Diagnostic X-Ray System with Digital Panel

DateNov 28, 2012
StatusTerminated
ReasonWhen the operator moves an image to another view then the electronic marker might be placed at the wrong anatomical position in the resulting image.
Root causeSoftware design
ActionPhilips Medicals Systems distributed an "URGENT Field Safety Notice" alerting customers of the defect on November 28, 2012. Customers are requested to manually rotate images by using the rotation tool.Customers are also recommended to check whether the automatic marker is set correctly, especially after a manual image rotation is performed. The marker can be corrected manually with the annotation tool.
Quantity220 systems
DistributionNationwide (AR, AZ, CA, DE, FL, GA, IA, IL, IN, KS, MD, MI, MN, NC, NM, NY, OH, OR, PA, SC, TX, VT, WI)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111725MOBILEDIAGNOST WDR
Sedecal., Sa., Jul 19, 2011