Recall Z-1337-2013 ยท JIX Calibrator, Multi-Analyte Mixture

Ortho-Clinical Diagnostics

VITROS Immunodiagnostic Products CK-MB Calibrators, REF 128 6293, IVD, Ortho-Clinical Diagnostics High Wycombe HP12 4DP, UK For use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System for the quantitative measurement of CK-MB in human serum and plasma (EDTA or heparin).

DateMay 17, 2013 (initiated Apr 3, 2013)
StatusTerminated
ReasonInternal testing demonstrated that opened-reconstituted VITROS CK-MB Calibrators stored at -20 degrees Celsius (-4 degrees Farenheit) generated invalid calibrations that were detected by positively biased quality control results.
Root causeProcess control
ActionOrtho Clinical Diagnostic sent a " Important Product Correction Notification" letter dated April 3, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were informed of the issue and advised they must discontinue storage of opened-reconstituted vials of VITROS CK-MB Calibrators frozen at -20 degrees Celsius (-4 degrees Farenheit). The same day, the Distributor letters were sent via US Postal Service Priority Mail or FedEx overnight courier to US Distributor consignees who were informed of the issue a
QuantityDomestic: 3019 units, Foreign: 905 units
DistributionWorldwide Distribution - Nationwide including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K973030VITROS IMMUNODIAGNOSTICS PRODUCTS/CK-MB REAGENT PACK(GEM.1300)/CK-MB CALIBRATORS (GEM.C300)
Johnson & Johnson Clinical Diagnostics, Inc., Sep 4, 1997