Recall Z-1188-2013 ยท JAK System, X-Ray, Tomography, Computed

GE Healthcare, LLC

GE, VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspeed VCT, Discovery CT750HD. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission.

DateApr 29, 2013 (initiated Jul 25, 2011)
StatusTerminated
ReasonFastener bolts on the heat exchanger were not properly torqued and part of the heat exchanger can detach from the gantry and hit the inside of the covers and other components of the gantry possibly resulting in the oil conduit breaking, spilling oil within the gantry and outside the system. This issue affects GE Healthcare VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspe
Root causeProcess control
ActionNo consignee letter was sent out. At customer site Field Engineer replaced heat exchanger, the tube and the pump. Field Engineer also verified torque on all bolts. Field action was initiated on 7/25/2011 and has been completed as of 3/29/2013.
Quantity166
DistributionWorldwide Distribution-USA (nationwide) including DC and the states of AZ, CA, CO, CT, FL,GA, IL, IN, KY, MA, MI, MS,MO, NH, NJ, NM, NY, NC, OK, OR, PA, SC, SD, TX, UT, VA, WA, and WI and the countries of ALBANIA ARGENTINA AUSTRALIA BELGIUM BRAZIL CANADA CHINA COLOMBIA COSTA RICA CZECH REPUBLIC DENM
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013561LIGHTSPEED 4.0 CT SCANNER SYSTEM
GE Medical Systems, Nov 8, 2001
K030420LIGHTSPEED 5.0 COMPUTED TOMOGRAPHY SYSTEM
GE Medical Systems, Mar 10, 2003
K040372LIGHTSPEED 7.0 CT SCANNER SYSTEM
Ge Medical Systems, LLC, Mar 1, 2004
K081105GE LIGHTSPEED CT750 HD (LIGHTSPEED 8.0)
GE Healthcare, May 9, 2008
K081496OPTIMA PET/CT 560
Ge Medical Systems, LLC, Jun 12, 2008