Recall Z-1212-2013 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare, LLC

Optima MR430s MRI Scanner; 5000-0002 is indicated to use as a magnetic resonance imaging device of the leg (excluding the thigh), knee, ankle, foot, elbow, forearm, wrist, and hand.

DateMay 1, 2013 (initiated Mar 28, 2013)
StatusTerminated
ReasonCoolant supply lines may leak liquid onto the floor resulting in a slippery surface and fall hazard.
Root causeProcess control
ActionGE Healthcare sent an Urgent Medical Device Correction letter dated April 23, 2013, to all affected customers. The letter informed customers of a potential safety issue due to leakage from gradient coolant lines. Consignees were instructed to monitor the floor for leakage and continue to use if leakage is not observed. If leakage is observed consignees are instructed to stop using the system and notify their local service representative. GE Healthcare will correct all affected scanners by providing an update at no cost. Customers with questions were instructed to contact their local service
Quantity81
DistributionWorldwide Distribution - USA including AL, CA, CO, FL, ID, IL, MA, NH, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI; Worldwide: Argentina, Australia, Austria, Brazil, Canada, Chile, Denmark, Egypt, Finland, Germany, India, Italy, Japan, Korea, Malaysia, Netherlands, Norway, Peru, Poland, Russia, S
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K1032381.5T OPTIMA MR430S
Ge Medical Systems, LLC, Feb 11, 2011