Recall Z-1244-2013 ยท MQF Catheter, Assisted Reproduction
Smiths Medical ASD, Inc.
Wallace Sure-Pro Single Stage Embryo Replacement Catheter Product Code: PP623 Sure-Pro Embryo Replacement Catheters are used for the introduction of embryos into the uterine cavity following in vitro fertilization. Sure-Pro Ultra Embryo Replacement Catheters have the added benefit of being visible under ultrasound. These products can be used in either a one- stage or two-stage procedure. In the
| Date | May 3, 2013 (initiated Mar 28, 2013) |
| Status | Terminated |
| Reason | Unable to pass catheter through outer sheath during embryo transfer |
| Root cause | Process control |
| Action | The firm, Smiths Medical, sent an "URGENT FIELD SAFETY NOTICE" dated March 28, 2013 to its customers. The notification describes the product, problem and actions to be taken. The customers were provided with two options for proceeding: 1) continue to use product if the product passes a pre-test; or 2) return affected product to Smiths Medical for replacement or credit. In addition, the customers were instructed to inspect their inventory and quarantine all affected products; perform a count of affected products currently in inventory; complete and return the attached Confirmation Form via fax |
| Quantity | 800 units |
| Distribution | Worldwide distribution: USA (Nationwide) and countries of: Arab Emirates (AE) Ireland (IE), Albania, Israel (IL), Argentina (AR),Italy (IT) Armenia (AM) , Japan (JP), Austria (AT) Jordan (JO), Bahrain (BH) Kuwait (KW) Belgium (BE), Lebanon (LB), Belarus , Netherlands (NL), Brazil (BR),Nigeria (NG), |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K033084 | WALLACE (SURE VIEW) EMBRYO REPLACEMENT CATHETER AND TRIAL TRANSFER CATHETER Portex , Ltd., Dec 19, 2003 |
| K990349 | WALLACE MALLEABLE STYLET-18CM, MODEL 1816ST, WALLACE MALLEABLE STYLET-23CM, MODEL 1816NST CooperSurgical, Inc., Jul 26, 1999 |
| K990350 | WALLACE CATHETERS-18CM, MODEL 1816; 23CM, MODEL 1816N; 23CM, MODEL 2316 CooperSurgical, Inc., Jul 26, 1999 |