Recall Z-0620-2014 ยท JQP Calculator/Data Processing Module, For Clinical Use
GE Healthcare It
The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities associated with the provision and utilization of clinical laboratory services and facilities, e.g., the storing and delivering of analytical results. It is a specially designed data program application (software), which is supplied for installation in existing mai
| Date | Jan 2, 2014 (initiated Mar 12, 2013) |
| Status | Terminated |
| Reason | GE Healthcare is aware of a potential safety issue with the use of GE Centricity Laboratory Instrument Interface (IF) where free-text sent from IM is not being transcribed into free-text in iNET. The instrument may upload results to IF/iNET when order/request is not yet in IF/iNET. In such a situation, test level supporting free-text sent from the instrument is discarded (not captured). |
| Root cause | Device Design |
| Action | GE Healthcare sent an URGENT MEDICAL DEVICE CORRECTION letter dated March 12, 2013 to all affected customers. The letter notified customers of the problem, suggested that customers experiencing this problem should retransmit the result from the instrument once the order exists in iNET and verify that the item level free-text is present in iNET, and to contact their GE service representative to request or install the updated software. Customers with additional questions can contact the GE Healthcare IT Help Desk at 1-847-589-8496 or tgg.nahelpdesk@med.ge.com. For questions regarding this rec |
| Quantity | 5 units |
| Distribution | Worldwide Distribution - US including MN and Internationally to India, Australia, Canada, and England. |
| FDA record | Open on fda.gov |