Recall Z-0337-2014 ยท HGM System, Monitoring, Perinatal

GE Healthcare It

Centricity Perinatal and Centricity Intensive Care is intended to be used in clinical departments healthcare delivery systems such as a Labor and Delivery, Postpartum Maternal Care, newborn Nursery, critical care units including Neonatal Intensive Care Units (NICU), and may also be used in physicians offices and outpatient clinics. This product is primarily intended to serve the purpose of electro

DateNov 18, 2013 (initiated Mar 11, 2013)
StatusTerminated
Reason1) Under rare network conditions, a Centricity Perinatal process may maintain an incorrect count of Patient slots in use which may lead to an inability to access the patient roster. The user is temporarily unable to review or edit form and chart information. This could result in a potential delay of care; however active fetal strips and permanent patient data storage are not affected. 2) When
Root causeSoftware design
ActionGE Healthcare sent an URGENT MEDICAL DEVICE CORRECTION letter dated March 11, 2013, to all affected customers. The letters requested that direct accounts ensure that all potential users of the software were made aware of the potential problems and the recommended actions. Direct accounts were asked to return a customer response form and to return it to GE Healthcare via fax at 847-939-1522 or e-mail at QARAFMl1@ge.com. Version 6.94 of the software has been modified to correct these defects. GE Technical Support: 800-433-2009.
Quantity1,132 units
DistributionWorldwide Distribution - USA (nationwide) including Puerto Rico and the countries of: Australia, Canada, Denmark, Germany, Hong Kong, Korea, South Korea, Thailand and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111614CENTRICITY PERINATAL AND CENTRICITY INTENSIVE CARE
Ge Healthcare Integrated IT Solutions, Sep 26, 2011