Recall Z-1164-2013 ยท LLZ System, Image Processing, Radiological

GE Healthcare, LLC

GE Healthcare, Xeleris 2 Processing & Review Workstation. The display, processing, archiving, and communication of data acquired by Emission Tomography cameras used in diagnostic radiology, including procedures for planar imaging, whole body imaging, tomographic (SPECT) imaging, positron imaging by coincidence, attenuation correction, and anatomical image registration

DateApr 23, 2013 (initiated Oct 6, 2011)
StatusTerminated
ReasonGE Healthcare has become aware of a software issue in your Xeleris 2.1 workstation (that is used for processing Discovery 530/570 data), that may impact patient safety. On imported CT scans, Myocardial Perfusion Imaging (MPI) studies may create artifacts in attenuation corrected images. This may lead to a potential for misdiagnosis.
Root causeSoftware design
ActionGE Healthcare sent an Urgent Medical Device Correction letter dated October 5, 2011, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. Consignees were instructed to contact their local GE Healthcare Service Representative if they have questions or concerns. They may also contact GE Healthcare Call Center , United States or Canada at 800-437-1171. Other countries 262-896-2890. For questions regarding this recall call 262-513-4122.
Quantity11
DistributionWorldwide Distribution - USA including GA, and NY; and countries of: Canada, Israel, Italy and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051673XELERIS 2 PROCESSING AND REVIEW WORKSTATION
GE Medical Systems, Jul 18, 2005