ClearanceBrief
Recall Z-1150-2013 ยท FEB Accessories, Cleaning, For Endoscope

Advanced Sterilization Products

EVOTECH Endoscope Cleaner and Reprocessor System, P/N 50004. Designed to automatically clean and high-level disinfect flexible, submersible video or fiber-optic endoscopes.

DateApr 18, 2013 (initiated Mar 28, 2013)
StatusTerminated
ReasonAdvanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because they have received reports of smoke emanating from the system due to a circuit board malfunction that may occur in the Channel Control Manifold sub-assembly.
Root causeUnder Investigation by firm
ActionAn Urgent Medical Device Field Correction letter dated 3/28/13 was sent to all customers who purchased the EVOTECH Endoscope Cleaner and Reprocessor (ECR). The letter informed the customers of the problems identified and the actions to be taken. Customers are informed that the failure can potentially cause the following outcomes for the cycles listed below: 1. In the Full Reprocessing Cycle (Cleaning and HLD): there is a potential for residual rinse water and/or alcohol to remain in the endoscope channels after the cycle completes. 2. In the HLD-Only Cycle: there is potential for an ade
Quantity761 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082392EVOTECH ENDOSCOPE CLEANER AND REPROCESSOR
Advanced Sterilization Products, Nov 13, 2008