Recall Z-1286-2013 ยท JXI Cuff, Nerve

Integra LifeSciences Corporation

NeuraWrap Nerve Protector NeuraWrap Nerve Protector is indicated for the management of peripheral nerve injuries in which there has been no substantial loss of nerve tissue.

DateMay 10, 2013 (initiated Mar 28, 2013)
StatusTerminated
ReasonDue to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the specifications for the products.
Root causeProcess control
ActionThe firm, Integra, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated April 9, 2013, to all US, Canada, and non-EU Integra consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to verify if they have any lots of the listed product, if so, STOP using that product; remove the product from service; and promptly complete and return the PRODUCT RECALL RETURN ACKNOWLEDGMENT FORM via Email: FCA@integralife.com or FAX to : 1-609-275-9445. Customer Service department will contact you to return and replace the product. "However, if y
Quantity954 boxes
DistributionWorldwide distribution: USA (nationwide) including Puerto Rico and countries of: Arab Emirates, Argentina, Belgium, Brasil, Canada, Chile, China, Costa Rica, Denmark, Egypt, Spain, France, Great Britain, Italy, Japan, Peru, Singapore, Thailand, and South Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041620NEURA WRAP NERVE PROTECTOR
Integra Lifesciences Corp., Jul 16, 2004
K072207DURAGEN XS DURAL REGENERATION MATRIX
Integra Lifesciences Corp., Aug 23, 2007