Recall Z-1135-2013 ยท NPT Aortic Valve, Prosthesis, Percutaneously Delivered

Edwards Lifesciences, LLC

Edwards Lifesciences Crimpers, Model 9100CR26. The Crimpers are used to crimp the transcatheter heart valve onto the delivery system prior to use.

DateApr 17, 2013 (initiated Mar 26, 2013)
StatusTerminated
ReasonHoles in the sterile barrier pouch may be caused by handling of the device during manufacturing or once pouches have been removed from the unit box and stored on the shelf by customers.
Root causePackage design/selection
ActionEdwards Lifesciences notified customers via letter and instructed them to maintain the device in its protective unit carton until ready for use and to inspect the pouch for holes or tears prior to use as instructed in the product's labeling. If any holes or tears are found, customers are instructed to contact Edwards Customer Service to arrange for return and replacement of damaged product.
Quantity6,117 units
DistributionNationwide Distribution-USA (nationwide) including DC and the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.