Recall Z-1084-2013 ยท DZD Tube, Orthodontic
Ormco/Sybronendo
Bracket Buccal Tube U7L DB/UL22 Roth MT HK 0/-14/0 Product Usage: Bracket Buccal Tubes are intended to hold in orthodontic wires used to move teeth during orthodontic treatment
| Date | Apr 10, 2013 (initiated Mar 8, 2013) |
| Status | Terminated |
| Reason | Ormco Corporation is voluntarily recalling one (1) lot of Bracket Buccal Tubes as a result of a packaging error. More specifically, the Bracket Buccal Tubes product package correctly identifies the product as gingival (0) offset brackets, but actually contains distal (10) offset brackets. Use of the affected product may cause unintended rotation of the molars. Although reversible, a delay in |
| Root cause | Packaging |
| Action | Ormco sent an Urgent Medical Device Recall letter dated March 8, 2013 to all affected customers, via USPS 1st class mail. The letter identified the affected product, problem and actions to be taken. The consignees were instructed to complete the Return Form and return any affected product identified in stock. For questions contact Ormco Customer Care. |
| Quantity | 8210 |
| Distribution | Worldwide Distribution - USA (nationwide) including the country of EU. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.