Recall Z-1133-2013 ยท DQK Computer, Diagnostic, Programmable

GE Healthcare, LLC

GE Healthcare Muse v7 cardiology information system. Model number 2026443-001: MUSE 7.0 app cd, Model Number 2026443-015: MUSE 7.0.2 app cd, Model number 2026443-017: MUSE 7.1 app cd, Model number 2026443-029:MUSE 7.1.1 A app cd, and Model Number 2026443-031: MUSE 7.1.1 app cd. Intended to store, access and manage cardiovascular information on adult and pediatrics patients.

DateApr 17, 2013 (initiated Jan 4, 2012)
StatusTerminated
ReasonGE Healthcare conducted a recall on the Muse v7 because incorrect medications were displayed on the MUSE, which could result in incorrect treatment.
Root causeSoftware design
ActionConsignee letters were not sent for this recall. GE Healthcare's recall strategy was to update all affected systems that were distributed. Updating of the devices was carried out from February 2012 to August 2012 by GE Field personnel.
Quantity963
DistributionWorldwide Distribution -- USA, including all states except ME, and the U.S. Virgin Islands; and, the countries of AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CHILE, CHINA, CZECH REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MALAYSIA, NETH
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050614MUSE CARDIOLOGY INFORMATION SYSTEM
Ge Medical Systems Information Technologies, Apr 5, 2005
K072502MODIFICATION TO: MUSE CARDIOLOGY INFORMATION SYSTEM
Ge Medical Systems Information Technologies, Nov 19, 2007
K083639MUSE CARDIOLOGY INFORMATION SYSTEM WITH VMWARE
Ge Medical Systems Information Technologies, Feb 17, 2009