Recall Z-1055-2013 ยท KFX Assembly, Thigh/Knee/Shank/Ankle/Foot, External

LeMaitre Vascular, Inc.

Glow 'N Tell 20 cm Tape: 1100-00 (100 strips); 1100-50 (50 strips); 1100-20 (20 strips). Intended to be placed on the skin to assist during imaging procedures.

DateApr 4, 2013 (initiated Mar 18, 2013)
StatusTerminated
ReasonDevices were not sealed correctly during the manufacturing process, and the sterility of these products have been compromised.
Root causePackaging process control
ActionLeMaitre sent an "URGENT:VASCUTAPE RADIOPAQUE TAPE DEVICE FIELD SAFETY NOTICE" dated March 20, 2013 to affected customers. The recall letter contains a form that requests customers to return to LeMaitre Vascular, Inc. as a record of notification and reconciliation. The letter provides instructions on how the customer can inspect the defective tape seals and/or return the affected products for replacement/credit.
Quantity13560
DistributionWorldwide Distribution-USA (nationwide) and the countries of Canada, EU, and Asia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K932184STERILE LEMAITRE GLOW 'N TELL TAPE
Vascutech, Inc., Jul 22, 1993