ClearanceBrief
Recall Z-1123-2013 ยท FEB Accessories, Cleaning, For Endoscope

Advanced Sterilization Products

EVOTECH Endoscope Cleaner & Reprocessor System, P/N 50004 Designed to automatically clean and high-level disinfect flexible, submersible video or fiber-optic endoscopes.

DateApr 15, 2013 (initiated Mar 22, 2013)
StatusTerminated
ReasonAdvanced Sterilization Products (ASP) has determined that a limited number of EVOTECH ECRs manufactured between 2011 and 2012 did not receive sufficient High Potential (HIPOT) DC voltage testing.
Root causeNonconforming Material/Component
ActionAn Urgent Medical Device Field Correction letter, dated 3/22/13, was sent to customers who purchased the EVOTECH Endoscope Cleaner and Reprocessor (ECR) HIPOT Testing. The letter informs the customers of the problem identified and the actions to be taken. Customers are instructed on what to do if electrical shock occurs. Customers are instructed that they will be contacted in the next 2 weeks on behalf of ASP to schedule HIPOT testing. Customers with questions are instructed to contact Stericycle at (888) 965-5816. Customers are instructed to report complaints or suspected problems with t
Quantity273 units
DistributionNationwide Distribution-USA including DC and the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082392EVOTECH ENDOSCOPE CLEANER AND REPROCESSOR
Advanced Sterilization Products, Nov 13, 2008