Recall Z-1271-2013 ยท FPA Set, Administration, Intravascular

Baxter Healthcare Corp.

A) Product Code 2N8371: Non-DEHP Y-Type Catheter Extension Set Kit, 6.0", Vol. 1.0 mL, 2 Luer Activated Valves, Male Luer Lock Adapter, Single use only Product Usage: The IV extension tubing kits are intended to provide users with a ready to use set up for the administration and withdrawal of fluids avoiding the need to acquire/assemble components from various manufacturers. The IV extension tub

DateMay 7, 2013 (initiated Mar 28, 2013)
StatusTerminated
ReasonCustomers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties.
Root causePackaging process control
ActionBaxter sent an "Urgent Product Notification" dated March 28, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taked. Customers are advised to locate and remove all of the affected products from their inventory. If the products were further distributed, consigneees are requested to notify their customers of the recall. Products can be returned for credit by contacting Baxter. An acknowledgement form is required to be compelted by customers.
QuantityA) Product Code 2N8371: 513,000 units
DistributionWorldwide Distribution - USA Nationwide and the countries of Puerto Rico, Brazil, Canada and Columbia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112893CLEARLINK LUER ACTIVATED VALVE, CLEARLINK SYSTEM NON-DEHP CATHETER EXTENSION SETS
Baxter Healthcare Corporation, Oct 18, 2011