ClearanceBrief
Recall Z-1237-2013 ยท GCX Apparatus, Suction, Operating-Room, Wall Vacuum Powered

Stryker Endoscopy

Stryker Irrigator Reusable Tips: 3mm X 32cm Regular Tip, Non-Vented, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.

DateMay 3, 2013 (initiated Mar 27, 2013)
StatusTerminated
ReasonThe pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123C , instead of the correct 132-133C.
Root causeError in labeling
ActionStryker sent "Urgent Device Correction" letters dated March 27, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.
Quantity147,921 - all sizes
DistributionWorldwide Distribution: USA (nationwide) and countries of: Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, Phillipines, Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brasil, Romania, South Africa, Chile, Korea, China.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K954726STRYKER STRYKEFLOW SUCTION IRRIGATOR (MODIFICATION)
Stryker Endoscopy, Nov 29, 1995